FOM Name: Spectrometer UV VIS Lambda 10
Model: Perkin Elmer Lambda 10
Contact: Anna Kiyanova (anna.kiyanova@wisc.edu, 608-263-1735)
Center: SMCL
Location: B46 Engineering Hall
Lambda 10 UV/Vis spectrophotometer is a general purpose instrument for absorption spectroscopy in the ultraviolet and visible spectral regions. Besides the standard scanning spectra mode, the supporting software has capabilities to perform kinetic measurements as well as certain automatic data conversions(to concentration units, for example).
This is an accordion element with a series of buttons that open and close related content panels.
Configuration
- Optical scheme: scanning, single-beam
- Wavelength range: 190 nm to 1100 nm
- Wavelength accuracy: ±0.3 nm
- Wavelength reproducibility: ±0.1 nm
- Spectral bandwidth: 2 nm (fixed slit)
- Photometric range: Absorbance -6 to +6, Transmission 0% to 100%
- Photometric accuracy: Absorbance ±0.003 at 440 nm and Absorbance = 1
- Baseline stability: Absorbance < 0.0003 per hour (440 nm, after warmup)
- Scan speed: 30, 240, 960, and 2880 nm/min
- Cell holder: for standard 12.5 mm square outside dimensions cuvette
- Beam center height: 15 mm above cell holder bottom
- Beam cross-section: ca. 1 mm x 9 mm (width x height) at focal point